ABB MEASUREMENT & ANALYTICS | OPERATING INSTRUCTION | OI/AO2000-LS25-EN REV. E

From: | Author:kongjiang | Time :2025-06-18 | 761 Browse: | 🔊 Click to read aloud ❚❚ | Share:

General purpose and explosion-proof variants   Measurement made easy

Introduction

As an integrated part of the Advance Optima series the LS25 can be combined with further analyzer modules and completely operated remotely in Ethernet networks. The LS25 has ATEX, IECEx and CSA certificates for use in potentially explosive atmospheres. 

Safety

General information and instructions

These instructions are an important part of the product and must be retained for future reference. Installation, commissioning, and maintenance of the product may only be performed by trained specialist personnel who have been authorized by the plant operator accordingly. The specialist personnel must have read and understood the manual and must comply with its instructions. For additional information or if specific problems occur that are not discussed in these instructions, contact the manufacturer. The content of these instructions is neither part of nor an amendment to any previous or existing agreement, promise or legal relationship. Modifications and repairs to the product may only be performed if expressly permitted by these instructions. Information and symbols on the product must be observed. These may not be removed and must be fully legible at all times. The operating company must strictly observe the applicable national regulations relating to the installation, function testing, repair and maintenance of electrical products.

Warnings

The warnings in these instructions are structured as follows:

DANGER The signal word ‘DANGER’ indicates an imminent danger. Failure to observe this information will result in death or severe injury.

WARNING The signal word ‘WARNING’ indicates an imminent danger. Failure to observe this information may result in death or severe injury.

CAUTION The signal word ‘CAUTION’ indicates an imminent danger. Failure to observe this information may result in minor or moderate injury.

NOTICE The signal word ‘NOTICE’ indicates possible material damage.

Special operating conditions 

• The gas analyzer meets Ex approval only with a suitably approved purging and monitoring unit for the pressurized enclosure. 

• The maximum permissible purging gas 

pressure (see Purge parameters for pressurized enclosure on page 10) may not be up-scaled. If the purging and monitoring unit used does not have an integrated pressure reducer, the purging gas pressuremust be limited using an additional external pressure reducer. 

• The device contains a 3 V lithium battery. The housing must not be opened when a hazardous atmosphere may be present. The safety instructions in Opening and closing the housing on page 13 must be observed. 

• Possible danger of electrostatic discharge. The safety instructions in Operating instructions on page 21 must be observed.

Ambient conditions

Operating temperature −20 °C to +55 °C Storage temperature −20 °C to +55 °C


Commissioning Introduction 

Having completed the installation of the transmitter and receiver according to the previous sections the analyzer is ready for start-up.  

The start-up of the system consists of 3 main activities: 

1. Start-up of electronics. 

2. Alignment of transmitter. 

3. Alignment of receiver.

Switching on the power supply 

• Switch on the electronics.  

The LCD in the transmitter unit will enter start-up mode 

showing: 

The start-up sequence will make sure the laser is tuned to the correct temperature before it is switched on, and the instrument performs a self-test on all systems. Allow about 5 minutes for the instrument to start the laser. 

The LCD will probably show ‘Laser line-up error’ and ‘Low transmission’ when the laser has started up. This is normal and indicates misalignment of the receiver and transmitter units, i.e. the laser beam does not hit the detector inside the receiver unit. 

If you use a red laser alignment jig for alignment of the transmitter/receiver refer to Align analyzer with laser alignment device on page 52. 

Commissioning 

Align analyzer with laser alignment devicePrinciple of rough pre-alignment During pre-alignment, the purging flanges installed opposite to one another are aligned coaxially to one another using the laser alignment aid. 

Necessary tools and other equipment for the adjustment procedure: 

• 1 × laser alignment device (laser pointer and focusing 

screen) 

• 2 × spanners to adjust the flanges with (M 16) 

• 1 × Allen key (5 mm) for the stud screws。

Mounting the adjustment device 

1. Leave the device switched on and disassemble transmitter and receiver units from the relevant purging flange. 

2. Remove the adapter ring (item 2 in Figure 38) from the purging flange. 

3. Fix the laser pointer to the transmitter-side purging flange with the union nut. 

4. Fasten the focusing screen to the receiver-side purging flange with the union nut. Performing pre-alignment 1. Unscrew the 6 stud screws (item 6 in Figure 38) until the ends of the screws no longer stick out of the holes.   2. Switch on the laser pointer. 3. Align the purging flanges by adjusting the 4 screws (item 5 in Figure 38) until the laser beam hits the center of the focusing screen. 4. Screw in the 4 stud screws until they stop so that the setting is fixed. 5. Switch off the laser pointer. 6. Remove the laser pointer and focusing screen.

  • Vibro-Meter Rack VM600SYSMK2-ABE04x 6U 19-inch rack Mk2 system
  • Vibro-Meter Relay expansion RLC16 Mk2 SIL2 (single card) 600-044L
  • Vibro-Meter Relay expansion RLC16 Mk2 SIL2 600-04
  • Vibro-Meter Relay expansion RLC16 Mk2 600-045L
  • Vibro-Meter Relay expansion RLC16 Mk2 600-045
  • Vibro-Meter I/O interface card  IOCN Mk2    600-051
  • Vibro-Meter CPU controller   CPUM Mk2  600-050
  • Vibro-Meter I/O interface card    IOC10 Mk2 (rear card)    600-046
  • Vibro-Meter Temperature / Analog Acquisition  AMC10 Mk2  600-042
  • Vibro-Meter I/O interface card IOC4 Mk2 600-043
  • Vibro-Meter Protective measuring card MPC4 Mk2 SIL2 601-041-A10
  • Vibro-Meter Protective measuring card MPC4 Mk2 601-041-A0
  • Vibro-Meter PSU-VM600 VM600 Rack Power Module
  • Vibro-Meter ABE056 - 1U ultra-thin single-rack
  • Vibro-Meter ABE046 - 6U 19-inch rack
  • Vibro-Meter ABE045 - 6U 19-inch rack
  • Vibro-Meter ABE044 - 6U 19-inch rack (with expansion slots)
  • Vibro-Meter ABE042 6U 19-inch rack (4 slots)
  • Vibro-Meter I/O interface card XIO16T
  • Vibro-Meter XMC16 16-channel combustion monitoring card
  • Vibro-Meter XMV16 - 16-channel vibration condition monitoring card
  • Vibro-Meter RLC16 Relay Card
  • Vibro-Meter IRC4 Intelligent Relay Card
  • Vibro-Meter IOCR2 input/output rear interface card
  • Vibro-Meter CPUR2 Central Processing and Controller Card
  • Vibro-Meter I/O interface card  -  IOCR
  • VIBRO-METER CPU controller CPUR
  • VIBRO-METER IOCN 200-585-000-322 I/O interface card
  • VIBRO-METER IOCN 200-585-000-321 I/O Interface Card
  • VIBRO-METER 200-595-074-323 VM600 system CPU management card, non-redundant
  • VIBRO-METER CPUM 200-595-074-322 CPU Controller
  • VIBRO-METER CPU controller CPUM 200-595-070-321
  • Vibro-Meter IOC8T VM600 Mk1 I/O Interface Card
  • Vibro-Meter AMC8 Analog Monitoring Card
  • VIBRO-METER IOC4G2 I/O interface card
  • VIBRO-METER MPC4G2 Protective measuring card
  • VIBRO-METER IOC4T 200-560-101-017 VM600 Mk1 Isolated I/O Interface Card, equipped with MPC4
  • VIBRO-METER IOC4T 200-560-101-015 VM600 Mk1 interface card
  • VIBRO-METER IOC4T 200-560-000-114 I/O interface card
  • VIBRO-METER IOC4T 200-560-000-113 I/O interface card
  • VIBRO-METER IOC4T 200-560-000-111 Sensor wiring
  • VIBRO-METER MPC4 200-510-072-212 Protection Measurement Card
  • VIBRO-METER MPC4 200-510-071-114 Protective Measurement Card
  • VIBRO-METER 200-510-071-111 4-channel vibration protection + 2-speed channels
  • VIBRO-METER 200-510-070-111 4-channel vibration protection + 2-speed channels
  • ADLINK PICMG Single Board Computers NuPRO-E315
  • ADLINK High Speed Link (HSL) PCI-7853
  • ADLINK Digital I/O PCIe-7432
  • ADLINK Mini-ITX Embedded Boards AmITX-AL-I
  • ADLINK 6U CompactPCI 2.0 Blades cPCI-6860A
  • ADLINK NuPRO-900A PICMG Single Board Computers
  • ADLINK Analog Output PCI-6308 Series
  • ADLINK USB DAQ USB-2405
  • ADLINK CRR-LX800-E-500 Compact Industrial Computer Module (COM)
  • ADLINK 15-slot 6U PXI/CompactPCI Chassis with AC PXIS-3320
  • ADLINK NUPRO-842 PICMG Single Board Computers
  • ADLINK PCI-8124-C Advanced 4-CH Encoder Card
  • ADLINK PCI-8102 Centralized Motion Controllers PCI-8102
  • ADLINK PCIe-7200 High-Speed DIO
  • SEL SEL-7201 Feeder Protection Panel
  • SEL SEL-7206 Distribution Bus Differential Protection Panel
  • SEL SEL-4388 Mirrored Bits Tester
  • SEL SEL-T4287 Traveling-Wave Test System
  • SEL SEL-4000 Relay Test System
  • SEL SEL-9510 Control Switch Module
  • SEL SEL-9502 Contact Arc Suppressor
  • SEL SEL-9501 Contact Arc Suppressor
  • SEL SEL-2652 Trip Coil Monitor
  • SEL SEL-2894 Interface Converter
  • SEL SEL-2890 Ethernet Transceiver
  • SEL SEL-2886 EIA-232 to EIA-485 Interface Converter
  • SEL SEL-9220 Fiber-Optic Adapter for SEL-300 Series Relays
  • SEL SEL-2824 Multimode Fiber-Optic EIA-485 Transceivers
  • SEL SEL-2820 Multimode Fiber-Optic EIA-485 Transceivers
  • SEL SEL-2815 Fiber-Optic Transceiver/Modem
  • SEL SEL-2814 Fiber-Optic Transceivers With Hardware Flow Control
  • SEL SEL-2812 Fiber-Optic Transceivers With IRIG-B
  • SEL SEL-2810 Fiber-Optic Transceivers With IRIG-B
  • SEL SEL-2831 Single-Mode Fiber-Optic Transceiver/Modem
  • SEL SEL-2830 Single-Mode Fiber-Optic Transceiver/Modem
  • SEL SEL-2829 Single-Mode Fiber-Optic Transceiver/Modem
  • SEL SEL-C805 200 µm Multimode Fiber-Optic Cable
  • SEL SEL-C804 Multimode Arc-Flash Detection Fiber-Optic Cables
  • SEL SEL-9331 Power Supply
  • SEL SEL-9322 15 VDC Power Supply
  • SEL SEL-9321 Low-Voltage DC Power Supply
  • SEL SEL-LPS Linear Power Supply
  • SEL SEL-RPM Redundant Power Module
  • SEL SEL-7250 Digital Direct-Replacement Assemblies
  • SEL SEL-2533 Annunciator
  • SEL SEL-2523 Annunciator Panel
  • SEL SEL-2522 Alarm Panel
  • SEL SEL-2516 Rack-Mount Remote I/O Module
  • SEL SEL-2515 Remote I/O Module
  • SEL SEL-2507 High-Speed Remote I/O Module
  • SEL SEL-2506 Rack-Mount Remote I/O Module
  • SEL SEL-2505 Remote I/O Module
  • SEL SEL-3401 Digital Clock
  • SEL SEL-9929 Satellite-Synchronized Clock Display Kit
  • SEL SEL-3405 High-Accuracy IRIG-B Fiber-Optic Transceiver
  • SEL SEL-3400 IRIG-B Distribution Module
  • SEL SEL-2401 Satellite-Synchronized Cloc
  • SEL SEL-2407 Satellite-Synchronized Clock
  • SEL SEL-2488 Satellite-Synchronized Network Clock
  • SEL SEL-5051/5052 Client/Server Network Management System (NMS) Software
  • SEL SEL-5056 Network Management System (NMS)
  • SEL SEL-3031 Serial Radio Transceiverceiver
  • SEL SEL-3061 Cellular Router
  • SEL ICON Integrated Communications Optical Network
  • SEL SEL-3622 Security Gateway
  • SEL SEL-3620 Ethernet Security Gateway
  • SEL SEL-3620 Ethernet Security Gateway
  • SEL SEL-3610 Port Server
  • SEL SEL-2730U Unmanaged 24-Port Ethernet Switch
  • SEL SEL-2725 Five-Port Ethernet Switch
  • SEL SEL-2744 Ethernet Switch
  • SEL SEL-2743 Ethernet Switch
  • SEL SEL-2742 Ethernet Switch
  • SEL SEL-2741 Ethernet Switch
  • SEL SEL-2731 Ethernet Switch
  • SEL SEL-2902 RJ45 to DB-9 Adapter Panel
  • SEL SEL-3390S8 Serial Adapter Card
  • SEL SEL-3390E4 Network Adapter Card
  • SEL SEL-3390T Time and Ethernet Adapter Card
  • SEL SEL-3360 Computing Platform
  • SEL SEL-3355 Computing Platform
  • SEL SEL-5703 Synchrowave Monitoring
  • SEL SEL-5705 Synchrowave Reports